112
3) Press the ENTER key to run the calculation and display the confirmation
screen.
4) Press the OK key to accept the calculation, or the ESC key to cancel
and return to the CAL menu.
5) Press the MEAS key to return to the measurement screen.
Note: The new Load Factor and Offset values can be reviewed and
edited in the GEOM tabbed menu item.
6) Calibration is completed.