Instruction Manual

4.3 Turning The Power ON
Turn on the power by pressing the POW key. The current acoustic velocity
setting will be momentarily displayed prior to entering the Measurement
Mode (Mode H). The display should then read 0.000 inch (or 0.00 mm).
If the Good Coupling Indicator is flashing on and off, the red transmit
connector of the probe is not connected properly.
4.4 Turning The Power OFF
The Power will turn off automatically after three (3) minutes of non-use.
To manually turn off the power, press & hold the POW key for 3 sec.,
then release.
4.5 Zero Adjustment
This operation adjusts the zero point of the probe. Zero adjustment data is stored in
the instrument. It is recommended that the zero
adjustment data be refreshed once a day, preferably
before starting the day’s work.
1 . Apply couplant on the test plate surface and
place the probe on it.
2 . Make sure the probe is in good contact with
the test plate surface and press the ZERO
switch.
3 . The zero adjustment procedure takes place and the display reads 0.197" or
5.00 mm when the process has been successfully completed.
NOTE: When the sonic velocity is set to a value other than 5390 m/s, the
display reads 0.197" or 5.00 momemtarily when the ZERO switch
is pressed. Zero adjustment is nonetheless proceeding correctly.
If the material to be tested is signifcantly larger or smaller than
0.197" or 5.00mm, and the ZERO switch is pressed, the display
will read --- -and the zero adjustment process becomes invalid. In
this case, the zero adjustment should be made on a sample of known
thickness of the test material itself.
4.6 Using Couplant Fluid
Apply couplant fluid to the mesuring surfacae before measurement. The
couplant eliminates air between the probe and test surface, promoting the
transmission of the ultrasonic pulse.
NOTE: Never use organic solvents, including thinners and alcohols.
The surface must be cleaned of couplant after measurement.
TI-44N Operating Instruction Guide
6
8.0 TROUBLE-SHOOTING GUIDE
Description of Problem Possible Cause Action To Be Taken
Gauge will not power Battery voltage may be Replace with new AA
up too low battery
Gauge will not power • New battery may be • Check battery voltage
up even though battery too weak (should be 1.5 Volt)
has been replaced
• Back-up battery (lithium) • Return gauge for a new
could be too weak lithium battery
Good Coupling
• Probe or cable could be • Clean probe & surface
Indicator flashes on/off defective and retry measurement
• Foreign substance may • Return to manufacturer
be trapped between for inspection/repair
probe and sample
Good Coupling
• Insufficient couplant fluid • Add more couplant
Indicator not shown
• Paint or coating is too • Grind off paint and/or
thick or voids and/or corrosion
corrosion lies under paint
“......” is displayed Probe could be degraded Return to manufacturer
when ZERO key is or defective for inspection or repair
pressed
Displayed thickness Acrylic probe face is Carefully sand face of
value is unstable deteriorated probe using #500 paper
Can’t obtain measure- Pipe diameter (OD) may Consult factory
ment on small dia. pipe be less than 1”
(25 mm)
Couping indicator does • Insufficient amount of • Apply additional couplant
not appear while measure- couplant
ment is being taken Test material surface is • Grind/polish measuring
too rough surface to 50-100-s
smoothness
Test material bottom • Change measurement
surface is too rough point
Too much ultrasonic • Change measurement
attenuation due to point
material composition
Test material thickness is • Change measurement
outside the proper point
measuring range
Fluxuating readings Defective probe Return to manufacturer
TI-44N Operating Instruction Guide
15