Instruction Manual

– 10 –
4. Press the third button as long as the black bar stands
on the rst digit of the serial number. After a waiting
time of a few seconds the device accepts the entered
digit. Now the other 3 digits have to be entered.
When the 4-digit number has been entered correctly
(see picture), press the button to conrm.
5. The request Reinitialize? will appear at the display.
Hold the humimeter LM6 in one hand and make sure
that nothing stands behind the measuring eld (black
plate).
6. Now press the second button () with one nger. The
calibration procedure will take a few seconds. During
that time, the instrument has to be held up in the air
and the measuring eld must NOT be touched.
7. The device is ready when the shown picture on the
right disappears automatically.