4.0 FUNCTION MODES
To access and modify the function modes, follow this procedure:
1. Make sure that the gauge is powered off
2. Press and hold the ZERO key.
3. While continuing to press and hold the ZERO key, Press and release the
POWER key.
4. Release ZERO after the display
shows “f01.”
5. Press UNIT to to change the setting
within the function selected.
OR
6. Press PEAK to save the current setting and move to the next Function Mode
IMPORTANT NOTE: To scroll through the Function Modes, press the PEAK
key repeatedly. The modes will appear in the order shown in the chart below.
OR
7. Press ZERO to save and exit the Function Modes.
MEM
MEM
Setting Unit Options Default
Measurement Polarity f01 -0.001 (minus), 0001 (plus) 0001
Display Update Time f02 1, 2, 3, 5, 10, 20, (times/second) 3
Auto Power Off f03 10 (10 minutes) oFF (always on) 10
RS232C Baud Rate f04 2400, 4800, 9600, 19200 bps 2400
Response Time f05 3, 20, 150 (msec) 3
Output Type f06 ovEr, Hi-Lo ovEr
4.1 Measurement Polarity (f01)
This function allows you to change whether compression is shown as a
positive or negative force. once the compression display has been changed,
the tension display will then read as the opposite of compression.
– 28 –
12.0 EXTERNAL DATA PORT
12.1 Pin Assignment
12.2 RS-232C Output
The RS-232C data connection allows control from external devices and data
transfers.
Interface
Baud rate* 2400, 4800, 9600, 19200 bps
Length of data bit 8 bit
Parity bit None
Length of stop bit 1 bit
Flow control None
*The baud rate is selectable through setting f04 (see section 4.4). The
default factory setting is 2400 bps. consult your equipment manual or
manufacturer for the correct baud rate. ASCII code, alpha numerics and car-
riage returns are used for RS-232C data transfer