User guide

7
Operation
Mode Operation Selection
Depending upon the type of application, the DT-5TXR/DT-5TS offers different modes of operation; these modes and
their functions are:
Select the mode of operation as follows:
1. Press and hold the MODE button for approximately 3 seconds; “0” will begin flashing until one of the five modes
(see table above) appears on the display.
2. Each time one of the lower buttons (NUMERIC, NEGATIVE SIGN/DECIMAL POINT, BACKSPACE) is pressed, the
display will scroll through these modes as indicated below:
3. Press the SET button to store the mode of operation and exit.
MODE NOITCNUF
-01- RPM or rate measurements
-02- Ela
p
sed time measurin
g
/ monitorin
g
-03- Process time measurement
-04- Flow rate measurement
(
used onl
y
in Ja
p
anese market
)
-99- Self-test
(
see “Self-Test” section underSet-U
p
)
Setting Mode Parameters
Modes -01-, -02- and -03- have several parameters that must be selected and programmed into the DT-5TXR/DT-5TS.
In addition, there are also function settings to further customize the DT-5TXR/DT-5TS.
NOTE: Because the flow rate measurement mode (-04-) is only used in the Japanese market, it is not covered in this
manual. Contact Shimpo for additional information if necessary.
NOTE: If a mistake is made entering any number, simply press the BACKSPACE button to erase the last digit on the
primary display.
Mode -01- (RPM or Rate Measurement)
The following parameters must be set in mode -01- so that the DT-5TXR/DT-5TS will display the correct readings:
DISPLAY MEANING
9999 - 1noitulover rep sesluP1 _ _ 0001 1 ppr
2 Sensing UMPR p to 99999 _ 01000 1000 RPM
00010001999999 10000.0stinu yalpsiD3
4 Update time 0.2, 0.5, 1.0, 2.0, 5.0, 10, 15, 30, 60 seconds
_
1.0
_
1 second
5 Display Hold” after input signal ceases 0.1150 seconds _ _ 006.0 6 seconds
zHk 20.0 ,001 ,03 ,01retlif tupnI6
_
10
_
10 kHz
FACTORY SETTING
RANGEDESCRIPTIONPARAMETER