User's Manual
Table Of Contents
- 1 Using this guide
- 2 Applicable Documents
- 3 About C-Scan®
- 4 Indications, Contraindications, Adverse Events, Warnings, Cautions
- 5 C-Scan® system
- 6 Instruction for use for the clinic team
- 6.1 Preparations before procedure
- 6.2 C-Scan® Cap ingestion procedure
- 6.3 During the procedure days (from C-Scan® Cap ingestion to excretion)
- 6.4 Radioactive source not properly concealed – malfunction notification instruction
- 6.5 Post procedure- after C-Scan® Cap excretion
- 6.6 Analysis & report reception
- 6.7 C-Scan® user interface notifications
- 7 Instruction for use to the subject
- 7.1 Permitted activities during the procedure:
- 7.2 Prohibited activities during the procedure:
- 7.3 C-Scan® procedure
- 7.4 Daily routine during the procedure
- 7.5 Post procedure- after C-Scan® Cap excretion:
- 7.6 When to contact the Clinic
- 7.7 Analysis & report reception
- 7.8 C-Scan® user interface notifications
- 8 C-Scan® View Installation Guide
- 9 Using C-Scan® View Application
- 9.1 C-Scan® View Application Login Screen
- 9.2 C-Scan View Updates
- 9.3 C-Scan® Track Data Download and Upload to Server
- 9.4 Add a Failed Procedure to Database
- 9.5 Managing your local files
- 9.6 C-Scan® View Editor
- 9.6.1 Editor introduction
- 9.6.2 Slice Profile
- 9.6.3 2D XRF/CMT Map
- 9.6.4 Structural Model
- 9.6.5 2D Map Timeline
- 9.6.6 Play Bar
- 9.6.7 General Buttons
- 9.6.8 Report on Case Failure
- 9.6.9 Estimated Colon Path
- 9.6.10 Bookmarks
- 9.6.11 Suspect Metrics
- 9.6.12 Suspect Wizard
- 9.6.13 Pre-analysis Summary
- 9.6.14 Report Preview
- 10 Technical Description
- 11 Troubleshooting- C-Scan® View
Document Title
C-Scan System User Reference Manual
Document No. DOC0001667 Revision 14 Date 22 May 2022
Page 35 of 90
Before using this document, check that you use latest revision (see Arena)
* Printed revisions are for Reference Only*
9 Using C-Scan
®
View Application
C-Scan
®
View is a proprietary application that reconstructs the raw data from the C-Scan
®
Track to a viewable format for review and analysis by the application users. It is installed by
Check-Cap on a PC supplied by Check-Cap.
Note: The C-Scan
®
View workstation is designed to be used only with the C-Scan
®
System.
Connecting any other devices or peripherals not defined in this document can cause it to
malfunction.
Note: Check the completeness of the power cord before booting the C-Scan
®
View.
The C-Scan
®
View software is a client/server-based application that enables procedure data
download from the C-Scan
®
Track, data analysis, and report generation.
The C-Scan
®
View is comprised of the following components:
PC with USB ports
Monitor, Keyboard, Mouse
USB-C cable
The C-Scan
®
View main functions are:
Load and display procedure information and data
Images Review - enables the user to view structural information of the colon wall
Produce procedure report
Store procedure results on server
The C-Scan
®
View software requires user Login.
Application users’ types and roles:
1. Loader- perform C-Scan
®
Track data download to workstation and upload to application
database cloud server.
2. Analyst- perform pre-analysis on procedure data and upload to application database
cloud server.
3. Supervisor- review and approve analyst’s pre-analysis. Can also perform pre-analysis
and upload to application database in cloud server.
4. Physician- review and approve supervisor’s pre-analysis. Perform final analysis, create
report and upload all to application database in cloud server.
5. Users admin- can create and manage application users and permissions.