4.6 ERASING THE MEASUREMENT MEMORY
To initialise the measurement memory, press simultaneously on MEM when switching On with the
rotary switch and do not release until it is erased. The MEM symbol lights up, the Init message is
continuously displayed (top right) for 3 seconds and the number of addresses available appears on the
digital display.
After 3 seconds Init ashes once and a beep indicates that the memory has been erased. The number
of addresses available then changesto 1920 which is the maximum capacity of the memory.
Note:
Before the beep you can release the MEM button at any time. In this case, it will not be erased.
4.7 ACCESS TO SEVERAL SPECIAL FUNCTIONS
As applicable, it may be useful to cancel the Auto Off
(1)
and beep functions
(2)
. To do this, press HOLD
(1)
, when switching On with the rotary switch, then press MIN MAX
(2)
before releasing the rst button.
This method allows you to access several special functions normally accessible only when switching
the instrument On.
5. MEASUREMENT MODE
Whatever the measurement mode, the sampling time is always 250µs. The table below summarizes the
measurement times of the various modes described in this chapter.
The time required for updating analogue measurements shown by the bargraph is always 20 ms.
Measurement mode Symbol Digital measurement
time
Normal measurement 400 ms
Recording
(of MIN, MAX and AVG) RECORD 400 ms
Smooth measurement SMOOTH 4 s
Recording in smoothed values RECORD SMOOTH 4 s
Peak measurement PEAK 100 ms
Recording in peak values RECORD PEAK 1 ms