User Manual

Technical characteristics
79
Resistance measurement
In Channel 1
Ranges (full scale)
Ohmmeter Resolution Measurement current
80 0.01 500µA
800 0,1 50µA
8k 1 50µA
80k 10 2µA
800k 100 2µA
8M 1000 50nA
32M 10k 50nA
Accuracy
±(0.5% + 25 D) from 10% to 100% of the scale
Open-circuit voltage
3V
Continuity measurement
In Channel 1
Beeper
<30 ±5
Measurement current
0,5mA
Beeper response
<10ms
Diode test
In Channel 1
Voltage
Open-circuit: + 3.3V
Accuracy
±(0.5% + 5 D)
Measurement current
0.6mA
Capacitance measurement
In Channel 1
Ranges
Capacitance meter Resolution Measurement current
5mF 1µF 500µA
500µF 0,1µF 500µA
50µF 0,01µF 500µA
5µF 1nF 50µA
500nF 100 pF 50µA
50nF 10 pF 2µA
5nF 1 pF 2µA
Accuracy
- in the 5nF range (measurement with a shielded lead):
from 500 pF to 1nF: ±(6% +10 D)
from >1nF to 2nF: ±(4% +10 D)
>2nF: ±(2% +10 D)
- in the other ranges: ±(2% +10 D) from 10% to 100% of full scale
Cancellation of series and parallel R
parallel R >10 k
Use the shortest possible leads.
Frequency measurement
from 20Hz to 200kHz on a square- and sine-wave signal
from 20Hz to 20kHz on a triangular signal
Accuracy: 0.2%
Power measurement
The power measuremenet is available only in AC, AC<5kHz, and AC <625 Hz.
active
± (2% +25 D) from 40 to 1kHz, filters inactive
reactive
± (4% +25 D) from 1 to 10kHz, filters inactive
apparent