12.17.3. Measurements (USB)
Diagnosis
Use this table to troubleshoot problems on a measurement :
Measurement
Description
Diagnosis
VHIGH
Measurement of the high level
Termination problem
Junction connection (oxidation, bad contact,
etc.)
Cable length not compliant with standard
Significant noise (check the cable route, ground
braid not connected, faulty chassis-ground, etc.)
…
VLOW
Measurement of the low level
Time Rise
Rise time between 10% and 90% of
the signal amplitude
Unsuitable or damaged cable
Termination impedance incorrectly positioned
Significant noise (check the cable route, ground
braid not connected, faulty chassis-ground, etc.)
…
Time Fall
Fall time between 90% and 10% of
the signal amplitude
TRise-TFall
Difference between the rise time at
10% 90% and the fall time at 90%
10%
Unsuitable or damaged cable (the rise and fall
times increase with the cable impedance)
Significant noise (check the cable route, ground
braid not connected, faulty chassis-ground, etc.)
…
Time Data
Measurement made using the bit time
total
Unsuitable or damaged cable
Termination impedance incorrectly positioned
Significant noise (check the cable route, ground
braid not connected, faulty chassis-ground, etc.)
…
Jitter
Measurement made using the bit time
total
Significant noise (check the cable route, ground
braid not connected, faulty chassis-ground, etc.)
…