Appendix - "Bus analysis" mode
143
12.15.3. Measurements (RS232)
Diagnosis
Use this table to troubleshoot problems on a measurement :
Measurement
Description
Diagnosis
VLevel High
Measurement of the signal high level
Termination problem
Cable length not compliant with standard
Chassis-ground disturbance problem
Significant noise (check the cable route, ground
braid not connected, faulty chassis-ground, etc.)
VLevel Low
Measurement of the signal low level
Time Data
Measurement made using the bit time
total
Significant noise (check the cable route, ground
braid not connected, faulty chassis-ground, etc.)
Cable length not compliant with standard
Unsuitable or damaged cable
Time Rise
Rise time between -3V and 3V
Cable length not compliant with standard
Unsuitable or damaged cable (the rise and fall
times increase with the cable impedance)
Time Fall
Fall time between 3V and -3V
Jitter
Measurement made using the bit time
total
Significant noise (check the cable route, ground
braid not connected, faulty chassis-ground, etc.)
Over+
Positive overshoot measurement
Unsuitable cable impedance
Bus termination problem (termination absent,
major overshoot)
Significant noise (check the cable route, ground
braid not connected, faulty chassis-ground, etc.)
Over-
Negative overshoot measurement