User Manual

Functional Description
33
4.1.9. PASS/FAIL function, from the screen
Opens the « Pass/Fail » menu
The Pass/Fail function is used to compare the evolution of the real-time signal to a mask. If
the real-time signal satisfies the predefined mask, the signal Passes; otherwise, the signal
Fails.
Activates/deactivates the Pass/Fail mode
Starts the analysis.
Acquisition counter
Selects the source for application of the mask and analysis.
Defines the width of the mask.
When a value is entered, the message "mask calculation in progress" appears, then the
new mask is displayed
Defines the height of the mask.
When a value is entered, the message "mask calculation in progress" appears, then the
new mask is displayed