User Manual
7
2.1.2. DISPLAY FUNCTIONS
Display of waveforms (voltages and currents).
Display of frequency bar chart (voltages and currents) (excluding neutral current).
Screen captures (12 maximum).
Transients function (for the C.A. 8333 only). Detection and recording of transients (up to 51) between user-defined start and
stop dates and times. Recording of 4 complete cycles (one before the triggering event and three after) in the 6 acquisition
channels.
Trend recording (data logging) function. 2GB memory with date-stamping and user-defined start and stop dates for record-
ing, with a maximum of 100 recordings. Display, in bar chart or curve form, of the means of many parameters vs. time, with
or without minima and maxima.
Alarm function (for the C.A. 8333 only). List of recorded alarms (up to 4,662) exceeding thresholds defined in the configuration
menu. User-defined alarm monitoring start and stop times – not more than two monitoring operations
2.1.3. CONFIGURATION FUNCTIONS
Date and time setting.
Screen brightness setting.
Choice of curve colours.
Choice of management of switching off of the screen.
Choice of display in night mode.
Choice of calculation methods (non-active quantities broken down or not, choice of the coefficients of calculation of the K
factor, choice of reference for the level of harmonics (for the C.A. 8333 only).
Choice of distribution system (single-phase, two-phase, three-phase with or without neutral) and of the connection method
(standard method or 2-element method).
Configuration of recording, alarms (for the C.A. 8333 only) and transients.
Erasure of data (total or partial).
Display of software and hardware version numbers.
Choice of language.
Display of current sensors detected not detected, not managed, simulated or impossible to simulate (2-element connection
method and 4-wire three-phase connection) . Setting of voltage and current ratio, of the transduction ratios and of the sensitivity.