14. GENERAL SPECIFICATIONS
14.1. ENVIRONMENTAL CONDITIONS
The following chart shows conditions relating to ambient temperature and humidity:
85
75
45
10
-20 0 20 26 35 42,5 50 70
°C
95
3 4
2
1
%RH
1 = Reference range.
2 = Range for use.
3 = Range for storage with batteries.
4 = Range for storage without batteries.
Caution: at temperatures above 40°C, the device must be powered by the battery alone OR by the mains power unit alone; use
of the device with both the battery AND the mains power unit is prohibited.
Altitude :
Use < 2 000 m
Storage < 10 000 m
Degree of pollution : 2.
Use indoors.
14.2. MECHANICAL CONDITIONS
Dimensions (L x W x H) 200 mm x 250 mm x 70 mm
Weight approximately 2 kg
Screen 118 mm x 90 mm, diagonal 148 mm
Tightness
IP53 per EN 60529 when the device is on its stand, with no lead connected, with the jack cover and the cap on the USB
connector in the closed position
IP20 at the level of the measurement terminals
IK08 per EN 62262
Fall 1 m as per IEC 61010-1
14.3. OVERVOLTAGE CATEGORIES PER IEC 61010-1
The device is compliant with IEC 61010-1, 600 V category IV or 1000 V category III.
the use of AmpFLEX™ or of MiniFLEX or C193 clamps keeps the “device + current sensor” system at 600 V category IV
or 1000 V category III.
the use of PAC93, J93, MN93, MN93A, or E3N clamps downgrades the “device + current sensor” system to 300 V cat-
egory IV or 600 V category III.
the use of the 5 A adapter unit downgrades the “device + current sensor” system to 150 V category IV or 300 V category
III.
Double insulation between the inputs/outputs and earth.
Double insulation between the voltage inputs, the power supply, and the other inputs/outputs.