User Manual

50
The Campaign on standby message is displayed until the start time is reached. It is then replaced by the message Campaign
running . When the stop time is reached, the Programming a Campaign screen returns with the key. You can then program
another campaign.
During an alarm campaign, only the stop date field can be modified. It is automatically highlighted in yellow.
8.3. DISPLAY OF THE LIST OF CAMPAIGNS
To display the list of campaigns performed, press the key. The List of Alarm Campaigns screen is displayed. The list can
contain up to 2 campaigns.
Figure 74: List of campaigns display screen
If the stop date of the campaign is in red, it means that it does not match the stop date initially programmed:
either because of a power supply problem (battery low or disconnection of the device supplied by mains only),
or because the memory is full.
8.4. DISPLAY OF LIST OF ALARMS
To select a campaign, move the cursor to it using the and keys. The selected field is bolded. Then validate with the
key.
The device then displays the alarms in list form.
Name of the campaign.
Start date and time of the campaign.
Stop date and time of the campaign.
Figure 75: Alarm list screen
If an alarm duration is displayed in red, it means that it was cut short:
because of a power supply problem (battery low),
or because of a manual stoppage of the campaign (press on ) or deliberate switching off of the device (press on the key).
or because the memory is full.
or because of a measurement error.
or because of an incompatibility between the quantity monitored and the configuration of the device (for example withdrawal
of a current sensor).
In the last two cases, the extremum is also displayed in red.
To return to the List of campaigns screen, press .
Alarm date and time.
Level of filling dedicated to the
alarm mode. The black part of the
bar corresponds to the fraction of
memory used.
Target of the alarm detected.
Type of alarm detected.
Extremum of the alarm detected
(minimum or maximum depending
on the programmed alarm direction).
Alarm duration.
The choice of filter is dynamic. It
depends on the connection chosen.