6.4.2. THE L1 PHASE-TO-PHASE VOLTAGE HARMONICS DISPLAY SCREEN
This information concerns the har-
monic pointed to by the cursor.
Uh 03: harmonic number.
%: level of harmonics with the fun-
damental RMS value as reference
(%f) or (for the C.A 8333 only) the
total RMS value as reference (%r).
V: RMS voltage of the harmonic in
question.
+000°: phase shift with respect to
the fundamental (order 1).
max – min: indicators of the maxi-
mum and minimum of the level of
harmonics or by pressing the
key.
THD: total harmonic distortion.
Ud: phase-to-phase RMS distortion
voltage.
The horizontal axis indicates the
orders of the harmonics. The level of
the harmonic is given as a percent-
age with respect to the fundamental
or (for the C.A 8333 only) to the total
RMS value.
Rang DC: DC component.
1 to 25: harmonics of order 1 to 25.
When the cursor exceeds order 25,
order 26 to 50 appears.
Harmonic selection cursor. Use the or
key to move the cursor.
Display of the 3 phases 3L, of L1,
L2 or L3. To select the display press
or .
Figure 49: Example of L1 phase-to-phase voltage harmonics display
Note: Filters L2 and L3 display the phase-to-phase voltage harmonics for phases 2 and 3, respectively. The screen is identical
to the one displayed for filter L1.
6.5. EXPERT MODE (FOR THE C.A 8333 ONLY)
The Expert mode is available with a three-phase connection only. It is used to display the influence of the harmonics on the
heating of the neutral and on rotating machines. To display expert mode press the or key of the keypad. The selection is
highlighted in yellow and the screen simultaneously displays the expert mode.
From this screen, two sub-menus are available:
V for three-phase set-ups with neutral or U for the three-phase set-up without neutral.
A for the expert mode in current.
Note : The breakdown into sequences shown here is valid only in the case of a balanced load.
6.5.1. THE PHASE-TO-NEUTRAL VOLTAGE EXPERT MODE DISPLAY SCREEN
For three-phase set-ups with neutral, the V sub-menu displays the influence of the harmonics of the phase-to-neutral voltage on
the heating of the neutral and on rotating machines.
Harmonics inducing a negative
sequence.
Harmonics inducing a zero se-
quence.
Harmonics inducing a positive se-
quence.
%: level of harmonics with the fun-
damental RMS value as reference
(%f) or the total RMS value as refer-
ence (%r).
Figure 50: The phase-to-neutral voltage expert mode screen (three-phase set-ups with neutral)
For three-phase set-ups without neutral, the U sub-menu displays the influence of the harmonics of the phase-to-phase voltage
on the heating of revolving machines.