User Manual

37
6.3.2. THE L1 APPARENT POWER HARMONICS DISPLAY SCREEN
This information concerns the har-
monic pointed to by the cursor.
S-h03: harmonic number.
%: level of harmonics with the
fundamental apparent power as
reference (%f) or (for the C.A 8333
only) the total apparent power as
reference (%r).
+045°: phase shift of the voltage
harmonic with respect to the current
harmonic for the order in question.
min–max: maximum and minimum
levels of the harmonic in question.
They are reset when the harmonic
number is changed or the
key is
pressed.
The horizontal axis indicates the
orders of the harmonics. The level of
the harmonic is given as a percent-
age of the fundamental apparent
power or (for the C.A 8333 only) the
total apparent power.
Rang DC: DC component.
1 to 25: harmonics of order 1 to 25.
When the cursor exceeds order 25,
order 26 to 50 appears.
Indicator of energy consumed
for this harmonic.
Harmonic selection cursor. Use the or
key to move the cursor.
Display of the 3 phases 3L, of L1,
L2 or L3. To select the display press
the or .
Figure 47: Example of L1 apparent power of harmonics display
Note: Filters L2 and L3 display the apparent power of the harmonics for phases 2 and 3, respectively. The screen is identical to
the one displayed for filter L1.
6.4. PHASE-TO-PHASE VOLTAGE
The U sub-menu is available for all connections except 2-wire single-phase. This sub-menu displays the harmonics of the phase-
to-phase voltage.
6.4.1. THE 3L PHASE-TO-PHASE VOLTAGE HARMONICS DISPLAY SCREEN
This information concerns the har-
monic pointed to by the cursor.
U-h03: harmonic number.
%: level of harmonics with the fun-
damental RMS value as reference
(%f) or (for the C.A 8333 only) the
total RMS value as reference (%r).
V: RMS voltage of the harmonic in
question.
+000°: phase shift with respect to
the fundamental (order 1).
The horizontal axis indicates the
orders of the harmonics. The level of
the harmonic is given as a percent-
age with respect to the fundamental
or (for the C.A 8333 only) to the total
RMS value.
Rang DC: DC component.
1 to 25: harmonics of order 1 to 25.
When the cursor exceeds order 25,
order 26 to 50 appears.
Harmonic selection cursor. Use the
or key to move the cursor.
Display of the 3 phases 3L, of L1,
L2, L3. To select the display press
or .
Figure 48: Example of 3L phase-to-phase voltage harmonics display