6.1.2. THE L1 PHASE VOLTAGE HARMONICS DISPLAY SCREEN
This information concerns the har-
monic pointed to by the cursor.
V-h03: harmonic number.
%: level of harmonics with the fun-
damental RMS value as reference
(%f) or (for the C.A 8333 only) the
total RMS value as reference (%r).
V: RMS voltage of the harmonic in
question.
-143°: phase shift with respect to the
fundamental (order 1).
max – min: maximum and minimum
levels of the harmonic in question.
They are reset when the harmonic
number is changed or the
key is
pressed.
THD: total harmonic distortion.
Vd: RMS distortion voltage.
The horizontal axis indicates the
orders of the harmonics. The level of
the harmonic is given as a percent-
age with respect to the fundamental
or (for the C.A 8333 only) to the total
RMS value.
DC: DC component.
1 to 25: harmonics of order 1 to 25.
When the cursor exceeds order 25,
order 26 to 50 appears.
Harmonic selection cursor. Use the or key
to move the cursor.
Display of the 3 phases 3L, of L1, L2,
L3 or (for the C.A 8333 only) of the
expert mode (three-phase connec-
tion only - see §6.5). To select the
display press or .
Figure 43: Example of display of harmonics of L1 phase-to-neutral voltage
Note: Filters L2 and L3 display the harmonics of the phase-to-neutral voltage for phases 2 and 3, respectively. The screen is
identical to the one displayed for filter L1.
6.2. CURRENT
The A sub-menu displays the harmonics of the current.
6.2.1. THE 3L CURRENT HARMONICS DISPLAY SCREEN
This information concerns the har-
monic pointed to by the cursor.
A-h05: harmonic number.
%: level of harmonics with the fun-
damental RMS value as reference
(%f) or (for the C.A 8333 only) the
total RMS value as reference (%r).
A: RMS current of the harmonic in
question.
+179°: phase shift with respect to
the fundamental (order 1).
The horizontal axis indicates the
orders of the harmonics. The level of
the harmonic is given as a percent-
age with respect to the fundamental
or (for the C.A 8333 only) to the total
RMS value.
Rang DC: DC component.
1 to 25: harmonics of order 1 to 25.
When the cursor exceeds order 25,
order 26 to 50 appears.
Harmonic selection cursor. Use the
or key to move the cursor.
Display of the 3 phases 3L, of L1, L2,
L3 or (for the C.A 8333 only) of the
expert mode (three-phase connec-
tion only - see §6.5). To select the
display press or .
Indicator of the presence of non-zero
harmonics of order higher than 25.
Figure 44: Example of 3L display of current harmonics