User Manual
21
Figure 23: The Choice of coefficients of calculation of the K factor menu
Use the arrow keys (,, , ) to fix the value of coefficients q and e:
q: exponential constant that depends on the type of winding and the frequency.
The value of q can range from 1.5 to 1.7. The value of 1.7 is suitable for transformers having round or square conductors, in
all types of winding. The value of 1.5 is suitable for those in which the low-voltage windings are in tape form.
e: ratio between the losses linked to eddy currents (at the fundamental frequency) and resistive losses (both evaluated at the
reference temperature). The value of e can range from 0.05 to 0.1.
The default values (q = 1.7 and e = 0.10) are suitable for most applications.
Then validate with the
key. The device returns to the Configuration menu.
Note: The modification is impossible if the device is recording and/or searching for alarms (for the C.A 8333 only).
4.5.3. CHOICE OF REFERENCE OF THE LEVEL OF HARMONICS OF THE PHASES
The %f-%r menu defines the reference for the level of harmonics of the phases.
4.5.2. CHOICE OF COEFFICIENTS OF CALCULATION OF THE K FACTOR
The FK menu defines the coefficients used for the calculation of the K factor.
Figure 24: The Choice of Reference for the Level of Harmonics menu
Use the arrow keys (,) to fix the reference for the level of harmonics:
%f: the reference is the value of the fundamental.
%r: the reference is the total value.
Then validate with the
key. The device returns to the Configuration menu.
In the case of the level of harmonics of the V-h, A-h, and U-h phases, the fundamental and total values are RMS values. In the
case of the level of harmonics of the S-h phases, the fundamental and total values are apparent power values.
Note: The modification is impossible if the device is recording and/or searching for alarms.










