User Manual

24
In continuity, in the 300 W, range, the instrument emits an
audible signal below 30 W.
3.3.3. DIODE
Particular reference conditions:
Zero voltage.
Diode without resistance or capacitance in parallel.
Diode voltage measured between 0.29 and 2V.
3.3.4. CAPACITANCE
Particular reference conditions:
Zero voltage.
Capacitance without resistance in parallel.
Range 3 nF * 30 nF * 300 nF 3 µF
Measurement
range
400 pF to
2.999 nF
3.00 to
29.99 nF
30.0 to
299.9 nF
0.300 to
2.999 µF
Resolution 0,001 nF 0,01 nF 0,1 nF 0,001 µF
Intrinsic uncer-
tainty
5% ± 10 pt 5% ± 5 pt
*: In these ranges, subtract the no-load value (typically 50pF)
from all readings.
Range 30 µF 300 µF 3 mF 30 mF
Measurement
range
3.00 to
29.99 µF
30.00 to
299.9 µF
0.300 to
2.999 mF
3.00 to
29.99 mF
Resolution 0,01 µF 0,1 µF 0,001 mF 0,01 mF
Intrinsic uncer-
tainty
5% ± 5 pt
The 3mF and 30mF ranges are not available in automatic
mode.
3.3.5. CURRENT
Particular reference conditions:
DC signal 1% in AC measurements.
Range 30 Aac 300 Aac
Measurement range 0.5 A to 29.99 A 30.00 A to 299.9 A
Resolution 10 mA 100 mA
Intrinsic uncertainty 3% ± 5 pt
3.3.6. NON-CONTACT VOLTAGE DETECTION (NCV)
The instrument detects the line voltage at 230VAC with respect
to ground, at 50 Hz and at a distance of less than 5cm.