User Manual
8
3. RESISTANCE MEASUREMENT
1) Connect the 2 cables to the 4 measurement terminals, then the 2 Kelvin clips to the object to be tested, which must not be live.
2) Set the switch to 400Ω - 10 mA.
3) Start the measurement by pressing
the START/STOP button.
If the measurement is too low, turn the
switch to the next lower range and restart
the measurement. Continue until the
display shows at least 3 digits.
4) Press the START/STOP button again to
stop the measurement...
...or disconnect one of
the 2 clips.