4.2.3 Frequency
Special reference conditions :
150 mV < U < 600 V
0,15 A < I < 10 A
When the switch is on the Hz or Volts position, the 300 Hz filter is not operational.
When the switch is in the V
LowZ
position, the 300 Hz filter is activated for Volts and frequency
Range
600 Hz
6 kHz
60 kHz
Specified measurement range
10 - 600.0 Hz
0.01 - 6.00 kHz
0.01 - 10 kHz
Uncertainty ()
0.1 % R + 2 D
0.1 % R + 2 D
0.1 % R + 2 D
Resolution
0.1 Hz
1 Hz
10 Hz
- Below 10 Hz, the value is forced to zero.
- If the detection level is insufficient, or if the current or voltage value is forced to zero, the freq. is not determined "----- ".
4.2.4 Resistance
Special reference conditions: the input (+, COM) must not have been overloaded following the accidental application of
voltage on the input terminals when the switch is on the or T ° position. If this is the case, the return to normal can take
about ten minutes.
Range
Specified
measurement range
Resolution
Uncertainty ()
Measurement
current
Voltage in open
circuit
600 Ω
0 600,0 Ω *
0,1 Ω
1 % R + 3 D
1 mA
< 5 V
6
0 6,000
0,001
1 % R + 2 D
120 μA
60
0 60,00
0,01
12 μA
600
0 600,0
0,1
1,2 μA
6
0 6,000
0,001
1,5 % R + 3 D
120 nA
60
0 60,00
0,01
3 % R + 5 D
30 nA
4.2.5 Audible continuity
Response time < 100 ms
Range
Resolution
Uncertainty ()
Voltage in open circuit
Meas. current
600 Ω
0.1 Ω
Beep triggered
< 30 Ω + 3 Ω
< 5 V
1 mA
4.2.6 Diode test
Range
Resolution
Uncertainty ()
Voltage in open circuit
Meas. current
6 V
1 mV
Beep triggered
< 40 mV + 10 mV
< 5 V
< 1,1 mA