User Manual
21
21
4.2.3 Frequency
Special reference conditions :
150 mV < U < 600 V
0,15 A < I < 10 A
When the switch is on the Hz or Volts position, the 300 Hz filter is not operational.
When the switch is in the V
LowZ
position, the 300 Hz filter is activated for Volts and frequency
Range
600 Hz
6 kHz
60 kHz
Specified measurement range
10 - 600.0 Hz
0.01 - 6.00 kHz
0.01 - 10 kHz
Uncertainty ()
0.1 % R + 2 D
0.1 % R + 2 D
0.1 % R + 2 D
Resolution
0.1 Hz
1 Hz
10 Hz
- Below 10 Hz, the value is forced to zero.
- If the detection level is insufficient, or if the current or voltage value is forced to zero, the freq. is not determined "----- ".
4.2.4 Resistance
Special reference conditions: the input (+, COM) must not have been overloaded following the accidental application of
voltage on the input terminals when the switch is on the Ω or T ° position. If this is the case, the return to normal can take
about ten minutes.
Range
Specified
measurement range
Resolution
Uncertainty ()
Measurement
current
Voltage in open
circuit
600 Ω
0 – 600,0 Ω *
0,1 Ω
1 % R + 3 D
≈ 1 mA
< 5 V
6 kΩ
0 – 6,000 kΩ
0,001 kΩ
1 % R + 2 D
≈ 120 μA
60 kΩ
0 – 60,00 kΩ
0,01 kΩ
≈ 12 μA
600 kΩ
0 – 600,0 kΩ
0,1 kΩ
≈ 1,2 μA
6 MΩ
0 – 6,000 MΩ
0,001 MΩ
1,5 % R + 3 D
≈ 120 nA
60 MΩ
0 – 60,00 MΩ
0,01 MΩ
3 % R + 5 D
≈ 30 nA
4.2.5 Audible continuity
Response time < 100 ms
Range
Resolution
Uncertainty ()
Voltage in open circuit
Meas. current
600 Ω
0.1 Ω
Beep triggered
< 30 Ω + 3 Ω
< 5 V
≈ 1 mA
4.2.6 Diode test
Range
Resolution
Uncertainty ()
Voltage in open circuit
Meas. current
6 V
1 mV
Beep triggered
< 40 mV + 10 mV
< 5 V
< 1,1 mA