8.3 Housing Assembly Removal
a. Remove all power to the Channel Master.
b. Remove the I/O cable and place the dummy plug on the I/O cable con-
nector (see “I/O Cable and Dummy Plug,” page 18).
c. Set the transducer assembly (transducer face down) on a soft pad.
d. Loosen (do not remove) the four titanium bolts (M6) bolts to vent the
system.
e. Once all four bolts have been loosened, remove the four bolts that attach
the housing flange to the transducer head assembly.
f. Carefully lift the housing assembly straight up and away from the trans-
ducer assembly until you can gain access to the connector jack on the
top PC board (see Figure 14). Use care; the plastic mating surfaces
scratch easily. Do not damage the mating surfaces.
g. Squeeze the sides of the internal I/O cable connector to release it from
the jack (see Figure 15, page 23). Set the housing assembly aside.
Figure 14. Removing the Housing
NOTE. The cable attached to the housing is only long enough to
disconnect the internal I/O cable. There is NOT enough cable to set the
housing down next to the transducer Assembly.