User manual
5 Volt power output
Parameter Min. Typ. Max. Unit
Typical output voltage
12,13
4,75 5,0 5,25 Volts
Voltage drop due to cable impedance
14
0,150 Volts
Command and response times
Single value
Min.
3,16
Typ.
3,16
Max.
3,16
Unit
Analog input
17
0,80 0,93 6,00 ms
Digital input
17
0,70 0,86 4,00 ms
Digital output
18
0,32 0,38 2,00 ms
Single frame
Min.
3,16
Typ.
3,16
Max.
3,16
Unit
Analog inputs only
17,19
0,72 0,90 4,00 ms
Digital inputs only
17,20
0,72 0,90 4,00 ms
Analog and digital inputs
21
0,72 0,90 4,00 ms
Maximum number of frames for buffered mode
Frame setup Maximum number of frames
1x analog input 4.095
4x analog input 1.023
Digital port 4.095
5x analog input + digital port + counter 585
Maximum sample rate - continuous DAQ
Frame setup Typ. Max.
3,16
Unit
1 analog input 100.000 Frames/s
2 analog inputs 50.000 76.000 Frames/s
3 analog inputs 33.500 59.000 Frames/s
4 analog inputs 24.000 49.000 Frames/s
Digital port 100.000 Frames/s
4 analog inputs + digital port 20.000 42.500 Frames/s
CEBO-STICK Reference Manual Cesys GmbH
[ v1.1 ] 24/28