also true if CEBO LC is unpowered.
4. Keep source impedance below given maximum value to ensure proper readings with default settling
times. For greater source impedances You may be required to increase settling times.
5. These values were calculated as follows:
f
total
= f
measure
+ f
calibration device (max)
fmeasure
=(x
d
-x
r
) / x
r
f: relative error
x
r
: value from calibration device
x
d
: mean of n measurements from "Device Under Test"
n=1000/1000/500/100 at range 10V / 1V / 100mV / 10mV
6. A sinusoidal voltage with a 10V amplitute was connected to one channel. Another channel was shorted
to GND and the cross talk was measured at all ranges. No other channels was measured.
20/63