Instructions for Use Designation of the unit: Single-user vacuum-cleaner unit Unit type: CERCON CLEAN 230VAC Unit No.
/15 Dear Customer: Thank you for the confidence in our work that you showed us by buying this vacuumcleaner unit. In order for you be able to enjoy your unit and its capabilities for many years, please take the time to read these Instructions for Use carefully, especially before using the unit for the first time. Contents 1 Safety 2 1.1 Proper use 2 1.2 Potential hazards 2 1.3 Authorized operators 2 1.4 Safety precautions at the deployment site 2 1.
/15 1 Safety 1.1 Proper use The CERCON CLEAN single-user vacuum-cleaning unit is used in dental laboratories for the removal of the dust generated when working with dustproducing machinery. It is not permitted to attempt to use this unit for the removal of liquids. This vacuum-cleaning unit is suitable for removing Category C dust as per the requirements of the German employers mutual insurance associations pursuant to DIN EN 60 335-269 (ZH 1/487).
/15 Danger The unit is not suitable for use in rooms in which special conditions prevail (e.g. a corrosive or explosive atmosphere). 1.5 Conventions used for the safety instructions in these Instructions for Use Note Indicates a practical hint to the user or other particularly useful information. Caution Indicates a usage or mode of operation whose non-observance may result in irregular operation, damage to the unit or other problems.
/15 dust collected. The large-area ultra-fine filter downstream from the singleuse filter bag is capable of holding back Category C dust 1. Fig. 1: View from front left 1 Casing 2 Cover 3 Clamp locks (2 ea.) 4 Panel 5 Mounting cover 6 Thumb screws (2 ea.
/15 Fig. 2: Section through unit, viewed from the left 10 Plastic connector for filter bag 24 Sealing surfaces, circumferential 11 Filter bag 25 Seal, circumferential 12 Basket 26 Air intake opening 15 Duct, 63 mm in diameter 27 Air intake chamber 16 Duct, 40 mm in diameter 28 M6 nuts (2 ea.) 17 Cover for filter bag 29 Ledge with longitudinal bores 20 Pre-filtration chamber 30 Electrical wiring central 21 Floor rollers 31 Ultra-fine filter 22 Screws for bottom space (9 ea.
/15 3. Connect the milling unit and the vacuum unit with the control cable included. 4. Connect the unit to mains (230 V/50 or 60 Hz). 5. Select the correct frequency using the frequency selector (13). 6. Turn on the mains switch (34). - The vacuum process will not begin until the milling unit is activated. 2.4 Electrically connecting the unit (see Fig. 3 on page 5 and Fig. 7 on page 10) 1. Connect the unit to mains (230V/50 or 60 Hz) using the mains cable included (socket (8)). 2.
/15 Thanks to the built-on power stabilizer, the preset input air volume is kept constant as the mains voltage changes (between 180 and 264 V). 3.3 Automatic turn-off and filter change indicator (see Fig. 6 on page 9). Note As soon as the filter bags (11) has been filled to the maximum permissible, an underpressure switch will actuate the automatic turn-off function, and the filter change indicator on the milling unit will be lit.
/15 Fig. 4: Removing the filter bag Caution The single-use filter bags must never be emptied and reused. Not only might this constitute a health hazard, but it would also impair the function of the unit. Caution Do not use pressurized air to remove any dust present in the filtration chamber (20). Do not allow any foreign matter to enter the air intake opening (26) because this might damage the vacuum motor. 5. Insert the new filter bag snugly and check the filtration module (see Section 4.2). 4.
/15 Fig. 5. Inserting the filter bag 4.3 2. Slide the new filter bag (11) in its basket (12) into the unit. The duct of the filter bag (10) will engage the duct (15) inside the filtration chamber (20) tightly. 3. Slide the plastic connector (10) tightly onto the housing duct (15) of the vacuum unit. 4. Place the cover (2) on the housing (1) and tighten using the clamp locks (3). Cleaning (replacing) the ultra-fine filter (see Figs. 1 and 2 on page 4 and Fig.
/15 3. Loosen the two nuts (28) and lift out the ledge (29) with the longitudinal holes. 4. Caution Remove the ultra-fine filter (31). No dust may be present on the sealing surfaces (24) or in the air intake chamber (32), as the dust may damage the vacuum motor (23). The air intake chamber (32) and sealing surfaces (24) must be cleaned by vacuuming. Do not use pressurized air. 5. Insert the cleaned or new ultra-fine filter (31). 6. Slide the ledge (29) onto the threaded pins from above. 7.
/15 5 Electrical protection The entire unit is electrically protected using two fine-wire fuses (main fuse) 6.3 A/T. The two fuses are located in the fuse drawer (40) inserted in the primary power socket housing (8) ist. Fig. 7: Electrical protection 7 Control cable socket 13 Frequency selector 50/60 Hz 8 Primary power socket 40 Fuse drawer with fuses 6,3 A/T (2 ea.
/15 6 Technical specifications Designation of the unit Single-user vacuum-cleaning unit Unit type CERCON CLEAN 230VAC Dimensions 420 mm wide (including mains cable relief), 390 mm deep (with vacuum duct), 540 mm high Voltage 200 + 240 Volt (50/60 Hz) Overvoltage category II Power requirements Max. 800 W Air throughput max.
/15 Normal wear and tear (brushes of the vacuum motor) or damage caused by improper operation are not covered by this warranty. To be able to offer you our comprehensive service, we ask you to return to us the warranty slip (inserted in the front of these Instructions for Use) by fax or mail. Fax +49 7181 73139 Designation of the unit Single-user vacuum -cleaner unit Duplicate Unit type Warranty Slip Unit No.
/15 EC Directive for Electrical Equipment Designed for Use Within Certain Voltage Limits (Low Voltage Directive) 73/23/EEC EC Electromagnetic Compatibility Directive 89/336/EEC The following harmonized standards were applied: DIN EN 292 Sicherheit von Maschinen. DIN EN 61 010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use DIN EN 55 014 Funk-Entstörung von elektrischen Betriebsmitteln und Anlagen.