User Manual
Add alarm recipients
Log in to the account. Click on ”Administration” > ”Alarm recipients”. in the left hand menu. Add the re-
quired information of the alarm recipient using the button (+ Add new alarm recipient). Test the alarm
by clicking on the mail box symbol of the recipient to which the test should be sent. NOTE! Check
that it arrives.
Other information
Stainless steel screws work well as measuring points in the measurement object (wooden structure).
Pins/nails may drop out of the material as it dries.
Avoid interference with the measurement signal by following these points:
• The MM611 bus cable and measuring cables should be placed at least 1 meter away from electrical
equipment or cables
• The measuring cables are sensitive to ESD/static electricity. Take measures to protect them
• A jagged/uneven waveform in a report indicates interference at the measurement point
The sensor should be mounted with the three half-rings on the cover facing upwards. If the sensor
is to be mounted against a metal surface, magnetic mount 7040 0060 (or other non-metallic spacer)
must be used to bring the sensor out at least 10 mm from the mounting surface. Otherwise, the metal
surface will interfere with the connection of the sensor and the measurement data will have difculty
reaching our servers.
Support
All support is handled by our distributors. See celsicom.se for more information. User manuals and
other support material are available at support.celsicom.se.