Operating instructions

Monthly Maintenance
Copyright © 2012 EVG 123
5
5
5) Calculate new value for “LightIntegratorFactor” by using the
following formula:
Current Factor
LightIntegratorFactor =
Current Lamp Power
x Real Lamp Power
Example (using values from above):
0.04
60mW/cm²
* 90mW/cm² = 0.06
6) Change the registry parameter to the calculated value and restart
IO_Serivce to activate the changes.
Example:
7) After restarting the IO_Serivce the monitor signal should show the
real value. (~90mW/cm2).
330 of 370