Operating instructions

Monthly Maintenance
Copyright © 2012 EVG 111
5
5
5.16.1 Adjust Water Flow
Figure 120 - Flow Meter
1) Close and open each flow meter to make sure that it does not stick
and that the cooling line is not clogged.
2) Set the flow rates according to the table below (or according to
specifications from the system installation):
System Diameter
(in mm)
Passive
cooling
Active top-side
cooling
Active bottom-
side cooling
EVG501 150 / 200
1l/min
N/A N/A
EVG510 150 / 200
1l/min
N/A
1l/min
All others 150 / 200
1l/min 1l/min 1l/min
All others 300
1l/min 2l/min 2l/min
External cooling station (150, 200 and 300mm): 1l/min.
Notice: Do not adjust the flow rate to a different value! An inappropriate flow
rate may cause substantial damage to certain components of the system!
5.16.2 Chiller
The following Chillers are used on the EVG5xx systems.
Neslab System II Liquid to Liquid Heat Exchanger
Neslab Merlin M75D
Thermoflex 2500
Check the temperature setting of the chiller. The recommended
temperature is 18°-20°C.
318 of 370