8) The display on the measuring tool must look as shown in below
(depends on the measuring tool):
If the display looks different press button “P1”.
If measuring tool as shown in item 2 is used press button “CLR” to
switch between the different sensors.
Figure 104
9) Open the “Jobs” –window in the system software. Point with the
mouse onto the Chamber –symbol and press left mouse button to
open the additional menu for the chamber. Go to “Piston” and
press button “Set Piston Force”.
Figure 105 - Load Cell Test - Set Piston Force