Operating instructions
Weekly Maintenance
Copyright © 2012 EVG 63
4
4
5) Open Cover and insert Calibration Tool in opening. (Second
person shout hold the Calibration Tool/sheet with soft tension.
Otherwise Calibration tool/sheet could be crumpled)
6) Close bond cover.
7) Adjust calibration sheet to marking 1.
4.6.1.1 Check correct adjustment
• Set the stack height to 500µm.
• Activate wafer bow and pull out the paper strip. It should be possible
to pull it out slightly.
• When setting stack height to 450µm, the paper should be clamped,
at 550µm it should be loosen.