Operating instructions

Weekly Maintenance
Copyright © 2012 EVG 57
4
4
4.5 Cover Clamps
Check the pneumatic connection of the cover clamps for cracks and
leakages.
Check that the sensors on the clamps are tightened.
Figure 53 - Clamps
Check that the four clamps on each chamber are closed and locked
well and if they open smooth again.
Check the inputs in the Low IO, Chamber x „GW XCO #1-C3“:
Status will become green when the input is present.
Endswitches are in series for:
Cover unlocked
Cover locked
Cover unclamped
Cover clamped
264 of 370