Operating instructions

Weekly Maintenance
Copyright © 2012 EVG 53
4
4
Without wafer bow equipped:
Set ring to its very lower position and proceed with measurement.
With piston in upper position, remove center aluminum plate (see this
manual page 9) and apply caliper as shown in the figure below. Set
caliper to 0 before measurement.
Figure 48 - Apply Caliper
Execute piston down (empty bond tool loaded) and check stroke.
Note: This different routine is necessary due to the higher hydraulic fluid
volume of the larger cylinder. The volume in the pressure converter limits the
hydraulic stroke of the cylinder to 10mm.
Therefore, the pneumatic “wafer bow/ hydraulic down” function has to be used
to bring the top-heater down to a maximum separation of 10mm, from there the
hydraulic can go ahead and apply force.
In fact this means that the stack height should never be set higher than stack-
thickness +10mm
Check schematics 2.1.2 for details.
260 of 370