Operating instructions

Weekly Maintenance
Copyright © 2012 EVG 31
4
4
4.3.1.3 Troubleshooting
Max. Pressure cannot be achieved or hydraulic pressure is
OFF
Inlet air-pressure too low: Check inlet pressure to be at 6bar
(600kPa) /87psi
Adjustment of V/P converter is off or the V/P-converter is bad:
Check pneumatic pressure with a gauge (see values in above table)
and call EVG Service
Stack height too low: Increase stack height and try again
Hydraulic system contains air: Fill/bleed the system (Filing /
Bleeding the hydraulic system)
Max. Pressure cannot be kept
When the hydraulic pressure get lost after ~8min: increase stack
height and try again
The maximal hydraulic pressure can be achieved but not kept:
increase stack height and try again.
Bleed the hydraulic system and try again.
If this does not solve the problem call EVG-service.
“Hydraulic Fail” Error message in the software
Check filling-level in the hydraulic-fluid reservoir.
Check “sensitivity” pot on the filling level sensor amplifier (call EVG
service)
238 of 370