Operating instructions

Daily Maintenance
Copyright © 2012 EVG 21
3
3
Insert the Measurement plate between the Flag and the Bondchuck:
Figure 17 - Insert Measurement Plate
1 Measurement plate
2 Flag
Lift off between measurement plate and flag should be reached at a
clamping force of: see next chapter
3.7.7 Clamping Force
The appropriate clamping force range depends on the wafers that are
clamped. The following table gives a guideline. Any deviation from
standard wafers (non-Si wafers, special shape of the wafer edge,
thinner wafers, defective edge…) requires an investigation of the
appropriate clamping force.
Bondchuck/Wafer inch Clamp force [N]
6” Si wafer; standard thickness 8 - 12 N
8” Si wafer; standard thickness 8 - 12 N
1
2
228 of 370