Operating instructions
Bond Module
32 Copyright © 2013 EVG
4.3.1.2 Wait Temperature
With the wait temperature command, the system waits until the
specified temperature is reached (on top and/or bottom heater).
According to the process command before, the system will heat or cool.
Therefore the temperature will pass the defined temperature upwards
(higher than) or downwards (lower than).
The Timeout defines the maximum allowed wait time until the system
stops the process and a timeout error message is displayed.
It is recommended to use the Timeout feature only in well-known and
often used processes. The command is not useful for R&D processes
as long as parameters are changed often. The risk that the system
stops the process because of a wrong Timeout is high.
Figure 25 - Wait Temperature
Value
Description
Heater Target
Select the target heater.
Mode
Enter the wait mode.
Temperature
Enter the target temperature.
Timeout
Activate timeout.
Deadline
Enter a timeout value. It defines the maximum allowed wait
time for the heater temperature. It will be reached until the
error recovery action is activated.
RecoveryAction
Choose the error recovery action.