Technical information
APPENDIX C
3/31/2008 Rev. 4.3 C - 5
Use the same procedure described under Operating Instructions – “Adding Consumable” to refill the
soda lime tube. Replace the filters (see Filters section below). Remember to gently tap the tube to
help settle the material. Keep all containers of fresh soda lime tightly sealed to prevent absorption of
ambient CO
2
.
Silica Gel
The silica gel should last from four to eight hours of operation under most circumstances. Actual
operating time will depend on ambient humidity levels. The silica gel supplied is a beaded, dust-
minimizing material of an indicating variety. The fresh material is dark blue and turns yellow as it
nears exhaustion. The relative humidity of the output of the CI-301AD will increase as the silica gel
becomes exhausted. When this happens, replace the silica gel.
Use the same procedure described under Operating Instructions-Adding Consumable to refill the
silica gel tube. Replace the cotton ball filters (see Filters section below). Remember to gently tap the
tube to help settle the material. Keep all containers of fresh silica gel tightly sealed to prevent
absorption of ambient moisture.
The used silica gel may be discarded or dried in an oven. If the material is dried, it should be placed
in a sealed container before cooling. Note that the indicator incorporated into the silica gel may not
function properly after this process, even though the silica gel itself may be dry.
Water
Water should be added if the level falls significantly. Change the water if it becomes cloudy or
contaminated in any way. Empty the water tube if the CI-301AD is not used for more than one or two
days.
CO2 Cartridge
Screw the CO
2
cartridge into the regulator. Set the regulator to 6 psi. The cartridge may seem to be
tight before it is punctured, so make sure to tighten until there is pressure indicated on the meter. The
regulator must be set to “LO” to get a reading on the meter.
Filters
The cotton ball filters located at each end of the soda lime and silica gel tubes become clogged with
fine particles and must be replaced. These filters should be replaced each time the chemicals are
changed.
It is extremely important that the filters be kept clean at all times. The first indication of a clogged
filter will be instability in the output of the CI-301AD. This may be followed by a reduction in the
available flow.