Technical information

APPENDIX A
3/31/2008 Rev. 4.3 A-4
OPERATING TEMPERATURE CONTROL MODULE
The CI-510CS Temperature Control Module is to be used with an Open System leaf chamber. Before
connecting the Temperature Control Attachment to your chamber, moisten the two surfaces
(improving thermal contact), which will be touching. You may do this by applying a few drops of
water to each surface. Align the two small screws on the Temperature Control Attachment with the
threaded holes on the chamber bottom and tighten, using the Allen wrench provided. (see Figure A-
7). The Infrared Leaf Temperature sensor should be inserted into the respective hole on the chamber
bottom.
Figure A-7. Illustration of Leaf Chamber, Allen wrench and Temperature Control Attachment.
A 12V battery supply outlet is provided for the CI-510CS. Remove all moisture on the Controller
panel. Plug the power connector to the jack marked 12V DC on the panel (Figure A-7).
WARNING: Only the provided power source can be used for the CI-510CS. Using other power
sources or power adapters may damage the CI-340 and the module. This will void the warranty.
Figure A-8. Configuration for the CI-510CS Temperature Control Module.