Technical data
7
2.3 Environmental Emission Information
1. The peak sound level generated by the autoclave is less than 70 dBA
with background noise of 60 dBA.
2. The total heat per hour transmitted by the autoclave is < 100 Wh for
all models.
2.4 Operating Conditions
This device is to be used for indoor use.
This autoclave is intended for NORMAL environment conditions as
follows:
♦ Altitude up to 2000m.
♦ Room temperature range 5ºC to 40ºC.
♦ Installation Category II.
♦ Pollution Degree 2.
♦ Maximum relative humidity 80% for temperature up to 31ºC
decreasing linearly to 50% relative humidity at 40ºC.
♦ Mains
supply
voltage
fluctuations
up
to
+/-10%
of
the
nominal voltage.
The sterilizer should be loaded only with autoclavable material.
Caution!
Waste water should be brought into the public net in accordance with
the local rules or requirements, i.e.
ONLY NON-HAZARDOUS LIQUIDS SHALL BE DISPOSED IN
PUBLIC SEWAGE!
2.5 Utilities
Utilities Unit Value
V-A 1ph, 120V – 16A,50/60 Hz
Power supply (as appropriate)
V-A 1ph, 230V – 16A,50/60 Hz
Attention:
The electrical net must be protected with a current leakage safety relay.
The electrical network must comply with local rules or regulations.
2.6 Construction
The main parts of the autoclave are made of materials as indicated
below:
♦ Chamber is electro-polish and built of stainless steel 316 L.
♦ Door is made of stainless steel CF8.
♦ Trays are made of stainless steel 304.
♦ Water reservoir is made of hard plastic material.
♦ Door handle is made of hard plastic material, which is safe to touch
and thermo-insulated.
♦ Covers are made of aluminum sheet, coated with Epoxy paint.