Technical data

52
Solution
19.1 Press the STOP key for 2 seconds to abort any program that may still be running.
20.1 Clean the air jet (see sec 9.4).
20.2 Check for proper voltage at the wall outlet.
21.1 Sterilization temperature set to high. Maximum setting 274ºF (134ºC).
22.1 The autoclave may be overloaded (see sec 7 Preparation for Sterilization).
22.2 Replace the HEPA filter (see sec 9.2).
23.1 Check water level in reservoir.
23.2 Check for an obstruction at the bottom of the reservoir blocking the pickup tube.
23.3 Clean the water sensor (see sec 9.10).
23.4 Try filling the machine manually using the Manual Inlet key (see sec 3).
24.1 Use Chamber Brite™ to clean the autoclave; very dirty machines may require two
cleanings. (see sec 9.9).
25.1 Remove the trays and rack and clean as per instructions in sec 9.1.2.
Problem
19 Keyboard does not respond.
20. Autoclave takes too long to heat up.
21. Loud noise coming from reservoir.
22. Drying takes too long.
23. Chamber not filling.
24. Chamber very dirty.
25. Trays and rack very dirty.
If the problem persists, contact your dealer or Tuttnauer USA for further assistance.