Technical data
49
Solution
1.1 Make sure the main switch is in the ‘On’ position,
(see front view drawing at the front of this manual).
1.2 Make sure the power cord is properly connected to the machine and the mains.
(See rear view drawing at the front of this manual).
1.3 Check the reset button on the cut-out thermostat, (see rear view drawing at the
front of this manual).
1.4 Make sure the circuit breaker has not tripped. Lift the circuit breaker lever.
2.1 Check the air jet (inside the water reservoir), (see sec. 9.4 Air jet Cleaning
Procedure).
2.2 Make sure the machine has the proper amount of sterilization load.
3.1 Clean the water level electrode inside the vessel, (see sec 9.10).
3.2 Check that the door is fully closed, the door gasket is seated and there is no
steam leakage. Replace the door gasket if necessary, (see sec. 9.5 Replacing the
Door Gasket).
3.3 Check the leveling of the machine, (see sec. 6, Installation).
4.1 Clean the water level electrode inside the vessel, (see sec 9.10).
4.2 Check that the door is fully closed, the door gasket is seated and there is no
steam leakage. Replace the door gasket, if necessary, (see sec. 9.5 Replacing the
Door Gasket).
4.3 Check the leveling of the machine, (see sec. 6, Installation).
5.1 See Low Water problem above.
5.2 The sterilization time has been set for too long of a period, allowing the chamber
to run dry.
10. TROUBLESHOOTING FOR THE OPERATOR
This troubleshooting chart enables the user to solve minor malfunctions, prior to requesting service.
Only technical personnel having proper qualifications and holding technical documentation (including a technician manual) and
adequate information are authorized to service the apparatus.
Problem
1. The machine is not responding.
2. ‘Low Heat’ is displayed.
3. ‘Low Water’ is displayed.
4. ‘Low Pres’ is displayed.
5. ‘Low Temp’ is displayed.