Technical data

35
8. OPERATING INSTRUCTIONS
It
is
important
to
clean
the
hole
of
the air jet, as described in sec. 9.4
before starting operation of the autoclave for the first time.
1. Remove water reservoir cover. Pour distilled water into the reservoir, through
the opening on top of the autoclave, until it reaches the base of the safety
valve holder, approximately 0.7 gallons (3 liters).
2. Plug the power cord into the back of the autoclave and into the wall outlet.
3. Turn on the rocker switch mounted on the bottom of the front panel.
4. If a printer is installed, set the clock with the proper date and time.
5. Press the required key to select the required program. The light indicator of
the selected program is illuminated; indicating the program has been selected.
The preset data of the program, sterilization temperature, time and dry time
are displayed.
6. Load the material to be sterilized into the chamber according to instructions in
sec. 7 (Preparation Before Sterilization) and close the door making sure the
‘Close Door’ indicator is illuminated.
NOTE:
Due to the inherent elasticity of the door gasket, the CLOSE DOOR
indicator light may be illuminated green before a complete seal is made
between the door and the chamber.
Therefore, in order to ensure that the door is fully sealed, when the green
light has been illuminated continue to tighten the door bolt until “hand
tight”. Do not overtighten the bolt as this may result in damage to the
gasket.
Should
the
autoclave
fail
to
reach
the sterilizing temperature/pressure,
always check first that the door is fully sealed. If not, tighten the door
bolt further, as described above, until completely sealed.
7. Press the START key to put the autoclave in operation.
WATER INLET is displayed until the correct volume of water is
automatically introduced.
The autoclave starts performing the sequence of operations. The actual
measured values of pressure and temperature are displayed continuously (and
printed with optional printer).
The indicator lights HEAT, STE, EXH and DRY are turned on and off as
each stage is started and completed.
NOTE:
It is possible to change parameters, only when the autoclave is not in
operation. In order to change the fixed preset parameters, proceed as
follows:
To increase or decrease the sterilization temperature, sterilization time or
dry time, follow instructions from sec. 3.2.2.
Pressing
the
UP/DN
key,
advances
the
setting
upward
or downward by
one
unit.
The
displays
are
updated
with
every
change
in
the
preset
data.
8. During any program that has a drying stage scheduled, the dry stage begins
after the steam exhaust stage. The autoclave is equipped with an air
compressor that, during the drying stage, draws air through a HEPA filter
(0.2µm), pushes that air through the heated chamber and out the air outlet
valve to remove moisture and facilitate the drying operation. Drying is
performed with the door closed.