Technical data

29
6.2 Water Filling
6.2.1 Automatic Water Filling
The proper amount of water for automatic filling in your new
Tuttnauer autoclave has been preset at the factory. However, if
in routine operation there is inadequate water in the chamber,
the operator can adjust the level with the automatic built-in
system by doing the following:
1. Press STOP key repeatedly until the message “code xxx”
appears on the display.
2. Use the UP or DN arrow keys to change the code to 105, then
press the STOP key.
3. A message will be displayed saying “Water in = xx sec”.
4. Press UP or DN keys to increase or decrease water inlet time.
5. Press the STOP key to enter new water inlet time into
memory.
6. If necessary, press the STOP key again to bring up the
program display.
Note:
DO NOT attempt to automatically fill the chamber with the DOOR
open. Water will overflow out of the chamber.
Warning:
If it becomes necessary to RESET the software program it will be
essential to repeat all steps listed above. This will ensure that the
correct amount of water enters the chamber for operation.
6.2.2 Checking the automatic fill
To check the automatic fill, follow these steps:
1. Remove any water that is in the chamber.
2. Make sure the unit is turned on.
3. Place a collecting vessel under the autoclave's door.
4. With the door open, press and hold the door switch, then
press the START key.
5. When water starts flowing into the chamber, release the door
switch.
6. Water should come beyond the groove at the front of the
chamber, and a small amount will pour into the collecting
vessel.
7. After the automatic filling is completed, tilt the autoclave and
pour all the water in the chamber into the collecting vessel.
8. Measure and verify that the amount of water pumped into the
chamber is 500-600 ml.
9. If the water is not filling correctly, follow the adjustment
procedure above.