PRINTER OUTPUT DESCRIPTION
Autoclave No: 01 Number of the autoclave with respect to other units in the
same location.
Load number: 0005 Load number. Useful to determine when to clean the
chamber. (Upon reaching 255, this number is reset to 0).
Operator :___________ To be filled in manually by operator.
O.K. Cycle completed successfully.
−−−−−−−−−−−−−−−−−−−
D20 220°F 00P The time, temperature and pressure during drying.
-----------------------
E20 251°F 02P The time, temperature and pressure during exhaust.
-----------------------
S20 273°F 31P The time, temperature and pressure during sterilization.
*
* Prints sterilization data every 1 minute.
*
S13 273°F 31P The time, temperature and pressure during sterilization.
S12 273°F 31P The time, temperature and pressure during sterilization.
----------------------
H08 231°F 10P The time, temperature and pressure during heating.
H04 137°F 00P The time, temperature and pressure during heating.
*
* Prints heat up data every 4 minute.
*
H00 72°F 00P The time, temperature and pressure during heating.
----------------------
MN TEMP PRES
DRY: 30min Drying time for selected program.
TIME: 08min Sterilization time for selected program.
TEMP: 273°F Sterilization temperature for selected program.
PROG: PKG Selected program: Unwrapped instruments.
TIME: 15:12:06 Time sterilization cycle begun.
DATE: 07:31:00 Date sterilization cycle begun.
Version:
T01EAWP
Number and version of the program.
Legend
MN Time elapsed in minutes
H Heating stage
S Sterilization stage
E Exhaust stage
D Drying stage
P psi