Technical data

23
4.2 PROGRAM 2. Wrapped Instruments and Porous Loads
For wrapped instruments and materials, when the manufacturer
recommends autoclaving at temperatures between 250°F and 274°F
(121°C and 134°C) with a drying stage.
Nominal parameters default settings
Sterilization temperature: 273°F (134°C).
Sterilization time: 7 minutes.
Dry time: 30 minutes.
Operations sequence:
Heating by actuation of electrical heaters until the sterilization
temperature is reached.
Sterilization temperature is maintained constant for the preset
sterilization time.
Fast exhaust, steam is exhausted out of the chamber at a fast rate until
pressure drops to 4 psi abs. (124 kPa abs.).
Dry heating and forced air circulation for 30 minutes to remove
leftover moisture from the instruments and wraps.
TIME
PRESSURE
(kpa)
TEMPERATURE
Ambient Pressure
and Temperature
t1 t2
t3
t1 = Steam generation stage
t2 = Sterilization stage
t3 = Fast exhaust stage
= Pressure
=
Temperature