Technical data

19
(13) STOP
This key issues the only command accepted by the system
during the running of a program. Pressing this key for over 1
second causes the program to immediately cease running and
enters the EXHAUST stage, at the end of which the MAN.
STOP” message will be displayed. This key has no function
when the system is not in operation and its only use is to
manually stop a cycle. In normal working conditions on
completion, the system automatically terminates the cycle,
without use of this key.
The STOP key does not function in EXH stage.
3.3 Description of the Display Panel
The display is comprised of 16 characters in one row and is divided into 4
sections.
The first section to the right, continuously shows the actual pressure within
the chamber. This happens whether the system is running a program or not
(provided the main power switch is turned on).
The three other sections are designated to show the parameters of the
selected program, or operating messages. When the system is running a
sterilization program, the sterilization temperature will be displayed above
the TEMP key. Sterilization time will be displayed above the STE. TIME
key and the drying time for the selected program will be displayed above
the DRY TIME key.
If the program is aborted as a result of parameters exceeding the controlled
limits (high, low pressure or temperature etc.) or a manual STOP
command, a message is displayed on the left side of the screen. When a
message is displayed, pressing any key erases the message and redisplays
the selected program screen.
When the system is running a program, the screen displays the current
temperature within the chamber and the remaining time for sterilization or
drying. The current real pressure inside the chamber is always displayed on
the screen.
3.4 Description of Displayed Messages and Safety Measures
Door Unlock Message is displayed and the DOOR CLOSED LED
indicator remains unlit, if the door is improperly closed
when the START button is pressed. If the door accidentally
opens during any stage of the cycle, the same message
appears, the DOOR CLOSED LED indicator will turn off,
and the system reacts as if the STOP key was pressed.
Man. Stop Message will be displayed and the FAIL indicator will light
after the STOP key is pressed for longer than 1 second.
Add Water Message is displayed and the respective red LED indicates
insufficient water in the water RESERVOIR. After water is
added to the reservoir, the START button must be pressed
again in order to start the required sterilization cycle.
Water Inlet During the automatic water fill, the message WATER
INLET is displayed as information to the operator.