Specifications

— 55 —
9. Pressing Top Menu button, the following display will appear.
*** < AD Revision Value Write > ***
< Max Level Write >
AD Value Read! 0xXXXX
< Min Level Write >
AD Value Read! 0xYYYY
Write Data Value
Max= 0xXXXX
Min= 0xYYYY
------ > Test END
Push 'OK' Key
10. Pressing OK button according to the message, Main Menu will appear.
Confirm that AD Code = O O O O O display appear in Main Menu.
11. After turning off the checking unit, install the battery pack again.
12. After performing the initialization, insert CF card installed the diagnostic program into the slot of the
checking unit again.
Doing so, Main Menu display will appear again.
8-3-9. Contrast Adjustment and Check
Important
The contrast adjustment can be performed using this check without rotating the variable resistor. This contrast
adjustment is similar to adjustment by rotating the variable resistor.
Therefore, there are no adjustments by rotating the variable resistor concerning BE-300.
Before performing this adjustment by the operation check, perform the following operation.
After tapping Settings in Top Menu, tap Brightness.
Doing so, the slider will exist in the position shown to below illustration.
If,nevertheless, the display contrast is abnormal, try to adjust the contrast using the following adjustment.
1. Tapping 11 Evol Write in Main Menu, the following display will appear.
2. Adjust the contrast pressing UP/ DOWN in the cursor button.
Doing so, above EvolVal value will change from 1 to 10.
Default value is 5.