Specifications
— 54 —
2. Supply DC 4.2 volt to the main battery terminal using DC stabilized power-supply instead of the main
battery.
To supply DC voltage, it is recommended to take advantage of the socket attached wires which is
shown with an arrow in the below photograph.
To take advantage of this part, wires must be cut off from the battery pack.
Built-in Battery Pack
This Battery Pack can be supplied one of the spare parts for BE-300.
Refer to PARTS LIST and EXPLODED VIEW in this manual for detail.
3. Perform the initialization pressing RESET button while holding down Power button.
4. After inserting the CF card installed the diagnostic program into the slot of the checking unit, Main
Menu will appears after about 5 seconds.
5. Tapping 9 VOLTAGE Write, the following display will appear.
*** < AD Revision Value Write >***
< Max Level Write >
Voltage Write - > TopMenu Key
Escape - > OK Key
6. Make the supply voltage DC 4.2 V precisely.
7. Pressing Top Menu button, the following display will appear.
***< AD Revision Value Write > ***
< Max Level Write >
AD Value Read! 0xXXXX
< Min Level Write >
Voltage Write - > TopMenu Key
Escape - > OK Key
8. Make the supply voltage DC 3.7 V precisely.