Specifications

R
&S
F
S-K40 Remote Control
Software Manual 1161.8209.42 - 04 81
The number of measurements for AVERage is defined with the command
SENSe:SWEep:COUN
t. It should be noted that synchronization to the end of the
indicated number of measurements is only possible in single sweep mode.
Example
DISP:TRAC1:MODE WRIT
'
Sets trace 1 to clear/write mode
Characteristics
*RST value: WRITe for trace 1, STATe OFF for trace 2 & 3
SCPI: conforming
DISPlay[:WINDow<1|2>]:TRACe<1…3>:Y[:SCALe]:RLEVel <numeric_value>
This command specifies the current internal instrument reference level used when
performing measurements.
The numeric suffix at WINDow<1|2> and TRACe<1 to 3> are irrelevant.
Example
DISP:TRAC:Y:RLEV?
'Returns the current reference level in use
Characteristics
*RST value: -
SCPI: device-specific
DISPlay[:WINDow<1|2>]:TRACe<1…3>:Y[:SCALe]:RLEVel:OFFSet
<numeric_value>
This remote control command specifies the external attenuation/gain applied to
measurements. The value corresponds to the reference level offset in spectrum
analyzer mode.
The numeric suffix at WINDow<1|2> and TRACe<1 to 3> are irrelevant.
Example
DISP:TRAC:Y:RLEV:OFFS 10
'E xte rn a l a tte nua ti on ( lev el off se t ) o f the a na lyz er is 10 dB
Characteristics
*RST value: 0 dB
SCPI: conforming
DISPlay[:WINDow<1|2>]:TRACe<1…3>:SMOothing:APERture <numeric_value>
This remote control command specifies the aperture of the window to be used
when trace smoothing is performed.
A single aperture applies to all traces which require smoothing and as such the
numeric suffix at WINDow<1|2> and TRACe<1 to 3> are irrelevant: