CONFigure:POWer:EXPected:RF <numeric_value>
This remote control command is used to specify the input power level of the source
signal as supplied to the Analyzer RF input.
Example
CONF:POW:EXP:RF 9
'The FS-K40 option assumes an input signal strength of 9 dBm
Characteristics
*RST value: 0
SCPI: device-specific
CONFigure:REFMeas ONCE
This remote control command is used to configure and initiate a reference
measurement.
Example
CONF:REFM ONCE
'The FSUP/FS-K40 option will start a reference measurement.
Characteristics
*RST value: -
SCPI: device-specific
3.5 DISPlay Subsystem
The DISPLay subsystem controls the selection and presentation of textual and graphic
information as well as of measurement data on the display.
DISPlay[:WINDow<1|2>]:TRACe<1…3>:MODE WRITe |MAXHold |MINHold |AVERage
|VIEW
DISPlay[:WINDow<1|2>]:TRACe<1…3>:Y[:SCALe]:RLEVel <numeric_value>
DISPlay[:WINDow<1|2>]:TRACe<1…3>:Y[:SCALe]:RLEVel:OFFSet <numeric_value>
DISPlay[:WINDow<1|2>]:TRACe<1…3>:Y[:SCALe]:AUTO
DISPlay[:WINDow<1|2>]:TRACe<1…3>:SMOothing:APERture <numeric_value>
DISPlay[:WINDow<1|2>]:TRACe<1…3>[:STATe] <Boolean>
DISPlay[:WINDow<1|2>]:TRACe<1…3>:Y[:SCALe] <numeric_value>
DISPlay[:WINDow<1|2>]:TRACe<1…3>:Y[:SCALe]:AUTO
DISPlay[:WINDow<1|2>]:TRACe<1…3>:MODE WRITe |MAXHold |MINHold
|AVERage |VIEW
This command defines the type of display and the evaluation of the traces in the
selected measurement window. WRITE corresponds to the Clr/Write mode of
manual operation. The trace is switched off (= BLANK in manual operation) with
DISP:WIND:TRAC:STAT OFF
.