Specifications
R
&S
F
S-K40 Remote Control
Software Manual 1161.8209.42 - 04 79
CALCulate<1|2>:SNOise<1 to 4>:X <numeric_value>
This command positions the selected spot noise marker to the indicated frequency
Example
CALC1:SNO:X 2MHz
'Positions spot noise marker 1 in screen A to time 2 MHz.
Characteristics
*RST value: -
SCPI: device-specific
CALCulate<1|2>:SNOise<1 to 4>:Y <numeric_value>
This command returns the measured spot noise marker result in the selected
measurement window. The units for this command are dBc/Hz.
Example
CALC1:SNO:Y?
'Outputs the measured value of spot noise marker 1 in screen A.
Characteristics
*RST value: -
SCPI: device-specific
3.4 CONFigure Subsystem
The CONFigure subsystem contains commands for configuring complex measurement
tasks. The CONFigure subsystem is closely linked to the functions of the FETCH
subsystem, where the measurement results of the measurements are queried.
CONFigure:POWer:AUTO <Boolean>
CONFigure:POWer:EXPected:RF <numeric_value>
CONFigure:REFMeas ONCE
CONFigure:POWer:AUTO <Boolean>
This remote control command is used to switch on or off automatic power level
detection. When switched on, power level detection is performed at the start of
each measurement sweep.
Example
CONF:POW:AUTO 1
'The FS-K40 option will automatically detect the input power
level
Characteristics
*RST value: 1
SCPI: device-specific